| Model Number |
220222904 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
| Date of Event |
12/21/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available for evaluation as it remains implanted in the patient at this time.If additional information becomes available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a fall where the patient sustained a medial malleolus fracture which is unrelated to the specific implant.The reason the physician is revising the case is due to a worry for implant loosening.The surgeon originally planned to explant the device and repair the fracture but the patient wanted it done in two stages.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a fall where the patient sustained a medial malleolus fracture which is unrelated to the specific implant.The reason the physician is revising the case is due to a worry for implant loosening.The surgeon originally planned to explant the device and repair the fracture but the patient wanted it done in two stages.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation.There are many clinical factors that can affect the results of any surgery, such as surgical technique, pre-operative and post-operative care, the implant, patient pathology and daily activity.Based on investigation, the root cause was attributed to a patient related issue.As the revision was caused, as reported, to a fall where the patient sustained a medial malleolus fracture and is unrelated to the specific implant.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|