MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V431M-0730
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device has been returned to olympus for evaluation and the investigation is in process.The olympus service center confirmed the reported event; the cutting wire was bent where it broke off at the proximal green marker and the knife length sticking out measured 25mm.No missing fragments.The coated portion covering the knife was missing and charred stain on the cutting knife due to thermal damage.The remaining wired passed the continuity inspection when checked with its active plug pin.The handle is functional and could extend and retract the remaining cutting wire inside the sheath smoothly.The ports and plug contact pin are intact.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Event or Problem Description
The customer reported to olympus the subject device cutting wire broke during an unknown procedure.The subject device was being used with a fuji endoscope and an erbe generator.The cutting wire appeared burnt where it broke and the catheter dislodged from the catheter.A non-olympus sphincterotome was used to complete the procedure.The customer reported no patient injury or harm.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the investigation results, a likely mechanism causing the broken cutting wire might be the following: -high frequency current was applied between the cutting wire and the tissue at the point of the contact.As a result, the current density at the contact area increased, and the cutting wire became instantly hot.-high frequency current was applied when the cutting wire and the tissue were being close to each other.As a result, an electrical discharge occurred, and the cutting wire became instantly hot.-high frequency current was applied when the distal end of the device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.The cutting wire and the tissue were being close to each other.As a result, an electrical discharge between the cutting wire and the distal end of the endoscope occurred, and the cutting wire became instantly hot.It is likely some kind of force might have applied to the coated portion of the cutting wire when the device was withdrawn from endoscope.This might have caused the coated portion of the cutting wire to detach from the cutting wire.As a result, the coated portion was partially missing.However, the exact cause of the reported event could not be identified.The following information is stated in the instructions for use which may have prevented the event: since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the wire is very short, the output is too high or activated while the wire touches metal parts of the endoscope, or the wire is tightened too strong.When the wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the wire will be pushed out toward the papilla or move sideways.If the wire breaks off, stop the output immediately and pull the slider completely to retract the broken wire into the tube.Then withdraw the instrument from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct and/or damage of the endoscope could result.Be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur.Do not activate output while the cutting wire touches the metal parts of the endoscope, or they are being close together.This could burn the tissue and/or damage the endoscope or the instrument.Olympus will continue to monitor field performance for this device.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional investigation results.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause of the torn coating could not be determined, likely factors causing the tear could have been the following: - the knife wire was deformed and the coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.- it is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.Although a definitive root cause of the damaged knife wire could not be determined, a likely cause of the cutting wire breakage might be the following reasons.- high frequency current was applied between the cutting wire and the tissue at the point of the contact.As a result, the current density at the contact area increased, and the cutting wire became instantly hot and melt.- high frequency current was applied when the cutting wire and the tissue were being close to each other.As a result, an electrical discharge occurred between the cutting wire and the tissue, and the cutting wire became instantly hot and melt.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu : - since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.- be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur.- when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.- do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.- if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Additional Manufacturer Narrative
This report is being submitted to correct the legal manufacturer¿s contact information and facility registration number.The facility registration number is (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
2-248-1 okkonoki, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13283117
Report Number8010047-2022-01572
Device Sequence Number4944692
Product Code KNS
UDI-Device Identifier04953170184192
UDI-Public04953170184192
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V431M-0730
Device Lot Number14V17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2021
Supplement Date Received by Manufacturer03/28/2022
01/12/2024
05/06/2024
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date04/26/2022
01/22/2024
05/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FUJI ENDOSCOPE AND AN ERBE GENERATOR.
Patient SexUnknown
-
-