MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 DOT PROV 4F DL FULL 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 DOT PROV 4F DL FULL 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a bent stylet is confirmed; however, the exact cause is unknown.One 3cg stylet was returned for evaluation.An initial visual observation showed use residue on the returned sample.A bend/kink was observed in the returned stylet approximately 5.7 cm proximal to the epoxy tip.No breaks were observed in the stylet and the distal tip was found to be present and intact.A microscopic observation revealed no breaks or tears in the polyimide tubing and the bend/kink in the polyimide tubing was found to be just proximal to the most proximal magnet within the polyimide.Insufficient evidence was found on the returned sample to indicate a specific root cause of the bend in the stylet; however, possible contributing factors include advancement of the stylet against resistance and damage during manipulation, handling, or use.A lot history review (lhr) of refs2667 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "during placement of another 4 fr.Dual lumen picc we had a wire with a significant bend in it upon removal from the patient.The wire was intact." (b)(6) 2022: returned stylet wire is kinked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROV 4F DL FULL 3CG
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13283248
Report Number3006260740-2022-00019
Device Sequence Number4872655
Product Code LJS
UDI-Device Identifier00801741130120
UDI-Public(01)00801741130120
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1274108
Device Lot NumberREFS2667
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/12/2022
Initial Report FDA Received Date01/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient SexFemale
Patient Weight55 KG
-
-