MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) displayed catheter not connected messages. No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to investigate to evaluate the unit. The fse tested the iabp unit with a balloon and system trainer, but was unable to reproduce the reported issue. However, the fse discovered four "check iab catheter" alarms in the logs to have occurred on (b)(6) 2021. The fse performed functional and safety checks to meet factory specifications. The iabp was then released to the customer and cleared for clinical service. A supplemental report will be submitted upon completion of our investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13285339
MDR Text Key287283045
Report Number2249723-2022-00113
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 01/18/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/28/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-