| Model Number |
9-VSDMPIHDE-024 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.The cause of the reported event could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported a 24 mm amplatzer post-infarct muscular vsd occluder was chosen for procedure, during procedure, the device was positioned in normal fashion.Prior to release the device presented in a cobra shape while in the right ventricle.The device was also determined too big via imaging and the device was removed.A smaller, 18mm amplatzer post-infarct muscular vsd occluder was chosen.The device was determined a fit and the device was released for placement.The patient remained stable throughout the procedure and has been discharged home.No additional information has been provided.
|
| |
|