U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH AIM-DEVICE F/GUIDEWIRE F/PFNA+TFN F/AP O; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH AIM-DEVICE F/GUIDEWIRE F/PFNA+TFN F/AP O; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.010.412
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that, the surface of the product had begun to crack.No patient involvement.This report is for one (1) aim-device f/guidewire f/pfna+tfn f/ap o.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.A device history record (dhr) review was conducted: visual inspection: the complaint device (aim-device f/guidewire f/pfna+tfn f/ap o) was returned for investigation.Additionally, a photo in the notes & attachments section was reviewed.It was identified cracked and broken sections of the epoxy covering across the middle section of the wire surface; the photo findings are consistent with physical investigation.No other issues were identified.Dimensional inspection: a dimensional inspection could not be performed as the device has post manufacturing damage and complex geometry; relevant dimensions cannot be measured.Document/specification review: the current drawings were reviewed.Investigation conclusion: the reported condition of the complaint device (aim-device f/guidewire f/pfna+tfn f/ap o) is confirmed.After investigation, the cracking condition is visible, but there are chipped details at the middle section of the device.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/ specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot - a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIM-DEVICE F/GUIDEWIRE F/PFNA+TFN F/AP O
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13285469
Report Number8030965-2022-00378
Device Sequence Number20060743
Product Code FZX
UDI-Device Identifier07611819349490
UDI-Public(01)07611819349490
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.412
Device Lot Number13F6968
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2022
Initial Report FDA Received Date01/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/10/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-