• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102208-500
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 12/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2021, senseonics was made aware of an incident where user experienced hypoglycemia due to inaccuracies in sensor readings.
 
Additional Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Additional Manufacturer Narrative
User did not want to provide any specifics of the event such as date, time and the glucose values.Due to lack of information, no investigation was possible for this complaint.The issue with sensor inaccuracies was addressed under (b)(4) where a sensor replacement was offered to the user.As part of resolution of (b)(4) for the issue with sensor inaccuracies, a sensor replacement was offered to the user.No separate resolution was possible for this complaint due to lack of information.H3.Device evaluated by manufacturer? no.H6.Investigation findings updated to 3221.H6.Investigation conclusions updated to 67.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Common Device Name
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key13285641
Report Number3009862700-2022-00007
Device Sequence Number4835127
Product Code QCD
UDI-Device Identifier00817491021441
UDI-Public00817491021441
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P160048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/18/2022
Device Model Number102208-500
Device Catalogue NumberFG-3300-01-001
Device Lot Number122945
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer12/21/2021
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/18/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age43 YR
Patient SexMale
-
-