| Model Number |
HS-004 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Infutronix is waiting for the iv set to be returned for evaluation.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, infutronix received a complaint from an end user: "an administration set model hs-004, lot 2008003 set broke off from the cassette." device operator was a registered nurse.Medication infused was 5fu.A patient was involved but not harmed.The contract manufacturer of the affected device(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A service provider of infutronix provided the evaluation report for the affected administration set on 03/18/2022.Visual inspection confirmed that the silicon-pvc adaptor on the proximal end of the cassette module was snapped and broken.The reported issue was confirmed.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr.
|
| |
|