INFUTRONIX, LLC. ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number HS-004
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Infutronix is waiting for the iv set to be returned for evaluation.
 
Event or Problem Description
On (b)(6) 2022, infutronix received a complaint from an end user: "an administration set model hs-004, lot 2008003 set broke off from the cassette." device operator was a registered nurse.Medication infused was 5fu.A patient was involved but not harmed.The contract manufacturer of the affected device(b)(4).
 
Additional Manufacturer Narrative
A service provider of infutronix provided the evaluation report for the affected administration set on 03/18/2022.Visual inspection confirmed that the silicon-pvc adaptor on the proximal end of the cassette module was snapped and broken.The reported issue was confirmed.
 
Event or Problem Description
This is a follow-up for the initially filed mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
INFUTRONIX, LLC.
177 pine street
natick MA 01760
Manufacturer (Section G)
INFUTRONIX, LLC
177 pine street
natick MA 01760
Manufacturer Contact
chaoyoung lee
177 pine street
natick, MA 01760
5086502007
MDR Report Key13286130
Report Number3011581906-2022-00001
Device Sequence Number4727521
Product Code FPA
UDI-Device Identifier00817170020031
UDI-Public00817170020031
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K153193
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2023
Device Model NumberHS-004
Device Catalogue NumberHS-004
Device Lot Number2008003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2022
Supplement Date Received by Manufacturer03/18/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-