• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 27MJ-501
Medical Device Problem Codes Obstruction of Flow (2423); Biocompatibility (2886)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 12/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
It was reported during a follow-up in the clinic the patient presented shortness of breath and chest pain.Upon further evaluation, it was discovered the patient had a "stuck" valve due to thrombus formation.The cause of the thrombus formation is due to the patients non-compliance with their prescribed anticoagulant therapy.The patient has also not had their inr followed as required.No additional information has been provided.On (b)(6) 2021, the patient underwent an explant procedure of the 27mm sjm masters series mechanical heart valve and it was replaced with a non-abbott device.The patient remained stable throughout the procedure.No additional information has been provided.
 
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
It was reported during a follow-up in the clinic the patient presented shortness of breath and chest pain.Upon further evaluation, it was discovered the patient had a "stuck" valve due to thrombus formation.The cause of the thrombus formation is due to the patients non-compliance with their prescribed anticoagulant therapy.The patient has also not had their inr followed as required.No additional information has been provided.On (b)(6) 2021, the patient underwent an explant procedure of the 27mm sjm masters series mechanical heart valve and it was replaced with a non-abbott device.The patient remained stable throughout the procedure.No additional information has been provided.
 
Additional Manufacturer Narrative
Additional information: g3,g6, h2, h6, h10.An event of the valve leaflets being stuck due to thrombus formation was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined, however it was noted that the patient was non-compliant with their anti-coagulation therapy and that their inr had not been followed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES MECHANICAL HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13286385
Report Number2648612-2022-00007
Device Sequence Number4599261
Product Code LWQ
UDI-Device Identifier05414734006460
UDI-Public05414734006460
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/06/2023
Device Model Number27MJ-501
Device Catalogue Number27MJ-501
Device Lot Number6311313
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer02/18/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date03/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-