| Model Number |
TMICL13.2 |
| Medical Device Problem Code |
Misfocusing (1401)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
12/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight: unk.Race: unk.Ethnicity: unk.Claim # (b)(6).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, -14.00/+1.0/074 (sphere/cylinder/axis), in the patients right eye (od) on (b)(6) 2021.The lens was explanted on (b)(6) 2021 due to patient experienced a refractive surprise (va was 20/400) and blurred vision.The lens was replaced with another same model/length but different diopter lens and the problem was resolved.The patient is doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned in liquid, in a vial.Visual inspection found one haptic torn.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3) device evaluation: the lens was returned in liquid in a vial.Visual inspection found haptic torn to the lens.Dimensional and functional inspection found the lens to be within specifications.Claim #(b)(4).
|
| |
|