| Model Number |
TMICL13.2 |
| Medical Device Problem Codes |
Misfocusing (1401); Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2; -6.50/+2.50/109 (sphere/cylinder/axis) implantable collamer lens into the patient's left eye (os) on (b)(6) 2021.The patient experienced excessive vaulting; blurry vision and the lens was exchanged on (b)(6) 2022 but this did not resolve the problem.It was later clarified that the patient still has fuzzy vision; pupils are dilated and prk surgery was performed.Patient still has blurry vision.Cause of the event was reported as "other- unexpected short cillary sulcus space".See mdr# 2023826-2022-00177 for replacement lens claim.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 device evaluation: the lens was returned in liquid in a specimen cup.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specifications.Claim # (b)(4).
|
| |
|