• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number TMICL13.2
Medical Device Problem Codes Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); Pupillary Block (2026); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 12/22/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2; -7.00/+3.50/100 (sphere/ cylinder/ axis) implantable collamer lens into the patient's right eye (od) on (b)(6) 2021.The patient experienced excessive vaulting; significant reduction of irido-corneal angles; angle closure with elevated iop; medication was prescribed (diamox); on (b)(6) 2021 the lens was exchanged with a shorter length lens but it was later reported that the issue was not resolved and the patient has pupil block/fixed pupil; pressure spikes and prk surgery was performed due to the vision.The cause of the event was reported as "other unexpected short ciliary sulcus space".See mdr # 2023826-2022-00176 for replacement lens claim.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
H3: device evaluation: the lens was returned in liquid in a vial.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specifications.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key13286583
Report Number2023826-2022-00174
Device Sequence Number17213602
Product Code QCB
UDI-Device Identifier00841542108645
UDI-Public00841542108645
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Model NumberTMICL13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Supplement Date Received by Manufacturer05/04/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
Patient SexFemale
-
-