| Model Number |
TMICL13.2 |
| Medical Device Problem Codes |
Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Pupillary Block (2026); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2; -7.00/+3.50/100 (sphere/ cylinder/ axis) implantable collamer lens into the patient's right eye (od) on (b)(6) 2021.The patient experienced excessive vaulting; significant reduction of irido-corneal angles; angle closure with elevated iop; medication was prescribed (diamox); on (b)(6) 2021 the lens was exchanged with a shorter length lens but it was later reported that the issue was not resolved and the patient has pupil block/fixed pupil; pressure spikes and prk surgery was performed due to the vision.The cause of the event was reported as "other unexpected short ciliary sulcus space".See mdr # 2023826-2022-00176 for replacement lens claim.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned in liquid in a vial.Visual inspection found no visible damage to lens.Dimensional inspection found the lens to be within specifications.Claim#: (b)(4).
|
| |
|