| Model Number |
115-222 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported right side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.
|
| |
|
Event or Problem Description
|
|
Patient reported right side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.A visual analysis of the returned device identified: underweight and opening.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: an opening sharp in the posterior side assessed as unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
Patient reported right side rupture.Device has been explanted.
|
| |
|