U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Flashers (1864); Halo (2227)
Date of Event 06/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that patient experienced "lights bursting" at night and halos immediately after the intraocular lens (iol) , model dfr00v 20.5 diopter, was implanted into the patient's ocular sinister (left eye).Account provided that patient was unable to drive at night and daily activities were significantly affected.The iol was explanted in a separate surgical procedure and a different model lens (dcb00) with the same diopter power was implanted as the replacement lens.There was no medical or surgical intervention required.Patient status was reported as fully recovered, no longer has complaints, and is happy with replacement lens.No further information was provided.
 
Additional Manufacturer Narrative
Product investigation was received, and the following fields were updated accordingly: section d-9: device available for evaluation: yes section d-9: date returned to manufacturer: 31-jan-2022 section h-3: device evaluated by manufacturer: yes device evaluation: visual inspection under magnification revealed that the lens was returned cut in pieces.Based on the return condition of the lens no further evaluation could be performed.No defects similar to or could have contributed to visual issue were observed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.No nc/er was found as part of this manufacturing records review.The lens diopter results obtained from the miq electronic system show that this production order (po) was released within diopter specifications.Historical data analysis: a search revealed that no other complaints have been received for this po number.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13287498
Report Number3012236936-2022-00152
Device Sequence Number14634589
Product Code HQL
UDI-Device Identifier05050474750685
UDI-Public(01)05050474750685(17)240308
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/08/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2021
Supplement Date Received by Manufacturer02/22/2022
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date02/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-