Model Number 102208-500
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 11/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This mdr is result of a retrospective review of complaints.The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event or Problem Description
On (b)(6) 2021,senseonics was made aware of an incident where user experienced hypoglycemia due to inaccuracies in sensor readings.
 
Additional Manufacturer Narrative
Based on the investigation analysis, it was confirmed the system asserted the hypo alerts from 4:10 pm est until 6:45 pm est on (b)(6) 2021.Per (b)(4), several attempts were made to contact the user to troubleshoot the issue where the user reported not hearing the alerts on the mobile device.Since the user was unresponsive to communication attempts, follow-up response is not available at the time to conclude if there was any issue with the mobile device or the transmitter.There could be no further investigation performed at this time.Per case notes, user mentioned that he has experience and knows how to treat the hypoglycemia episodes.No further resolution details are available for this complaint.H6.Investigation findings updated to 3221.H6.Investigation conclusions updated to 67.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Common Device Name
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key13287508
Report Number3009862700-2022-00013
Device Sequence Number4728090
Product Code QCD
UDI-Device Identifier00817491021441
UDI-Public00817491021441
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P160048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/19/2022
Device Model Number102208-500
Device Catalogue NumberFG-3300-01-001
Device Lot Number122946
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2021
Supplement Date Received by Manufacturer12/07/2021
Initial Report FDA Received Date01/18/2022
Supplement Report FDA Received Date01/20/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/19/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-