• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102208-500
Device Problem Imprecision (1307)
Patient Problem Hypoglycemia (1912)
Event Date 11/12/2021
Event Type  Injury  
Manufacturer Narrative
This mdr is result of a retrospective review of complaints.The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On (b)(6) 2021,senseonics was made aware of an incident where user experienced hypoglycemia due to inaccuracies in sensor readings.
 
Manufacturer Narrative
Based on the investigation analysis, it was confirmed the system asserted the hypo alerts from 4:10 pm est until 6:45 pm est on (b)(6) 2021.Per (b)(4), several attempts were made to contact the user to troubleshoot the issue where the user reported not hearing the alerts on the mobile device.Since the user was unresponsive to communication attempts, follow-up response is not available at the time to conclude if there was any issue with the mobile device or the transmitter.There could be no further investigation performed at this time.Per case notes, user mentioned that he has experience and knows how to treat the hypoglycemia episodes.No further resolution details are available for this complaint.H6.Investigation findings updated to 3221.H6.Investigation conclusions updated to 67.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key13287508
MDR Text Key285348446
Report Number3009862700-2022-00013
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491021441
UDI-Public00817491021441
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/19/2022
Device Model Number102208-500
Device Catalogue NumberFG-3300-01-001
Device Lot Number122946
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/07/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/19/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-