PENUMBRA, INC. BENCHMARK DELIVERY CATHETER; DQY Back to Search Results
Model Number BMK6F105M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a medical procedure using a benchmark 6f 071 delivery catheter (benchmark), a guidewire, and a 6f non-penumbra sheath.During the procedure, while advancing the benchmark over the guidewire and through the sheath, the distal portion of the benchmark broke off.Therefore, the physician used a snare device to remove the broken part of the benchmark.No additional information regarding the completion of the procedure was provided.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Additional Manufacturer Narrative
Please note that the device associated with this report was expected to be returned; however, additional information received from the penumbra regulatory consultant indicated that the device is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.Based on this information, corrections were made to: 1.Section d.Box 10.Device available for evaluation? (do not send to fda).2.Section h.Box 3.Reason for non-evaluation.3.Section h.Box 3.''other'' reason for non-evaluation.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Additional Manufacturer Narrative
Please note that the device in complaint was not expected to be returned; however, on 26-oct-2022 the device was received.Based on this information, corrections were made to: 1.Section d.Box 10: device available for evaluation? (do not send to fda) 2.Section h.Box 3: device returned to manufacturer)? 3.Section h.Box 3: device evaluated by manufacturer? additional information was added to: 1.Section d.Box 10: returned to manufacturer on (mm/dd/yyyy).Evaluation of the returned benchmark confirmed a fracture on the distal shaft.This damage was likely a result of the retraction against resistance reported during the procedure.The root cause of the resistance experienced during the procedure could not be determined.Further evaluation revealed the distal fractured segment was kinked and ovalized.This damage likely occurred during snaring the device to remove from the patient body and was incidental to the reported complaint.Evaluation also revealed a guidewire and a 5f select were stuck inside the benchmark.The benchmark was stretched, the catheter lumen was damaged, the internal coil winds were unraveled and wrapped around the 5f select and the guide wire.This damage was likely incidental to the complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13287591
Report Number3005168196-2022-00017
Device Sequence Number17245606
Product Code DQY
UDI-Device Identifier00814548014043
UDI-Public00814548014043
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K142321
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/30/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/16/2023
Device Model NumberBMK6F105M
Device Catalogue NumberBMK6F105M
Device Lot NumberF101198
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer05/11/2022
11/08/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date05/24/2022
11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
Date Report Sent to FDA01/01/2005
-
-