MEDOS INTERNATIONAL SàRL CH SYMPHONY OCT SYSTEM POLYAXIAL SCREW DIAMETER 4.0 ROD, 3.5X16MM 3.5 AND 4.0; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 1020-35-116S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 12/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent spinal reoperation due to broken symphony screw at c2 and reports of pain.The implant and fragments were successfully removed.The initial operation was performed on (b)(6) 2021.This report is for one (1) symphony oct system polyaxial screw diameter 4.0 rod, 3.5x16mm 3.5 and 4.0.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the product was returned to depuy synthes customer quality (cq) for evaluation.The depuy synthes cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the 4.0 ply scrw 3.5x16 was broken.Only the threaded shaft of the screw was returned.Images were also reviewed and confirmed the broken condition.No other issues were identified.The dimensional inspection was not performed due to post manufacturing damage.The observed condition of the 4.0 ply scrw 3.5x16 was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the 4.0 ply scrw 3.5x16 was observed to be broken.While no definitive root cause could be determined, it is probable that the 4.0 ply scrw 3.5x16 was broken due to unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMPHONY OCT SYSTEM POLYAXIAL SCREW DIAMETER 4.0 ROD, 3.5X16MM 3.5 AND 4.0
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle 02400
SZ   02400
6103142063
MDR Report Key13288595
Report Number1526439-2022-00131
Device Sequence Number11864520
Product Code KWP
UDI-Device Identifier10705034524382
UDI-Public10705034524382
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K192014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1020-35-116S
Device Catalogue Number102035116S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/23/2021
Supplement Date Received by Manufacturer01/28/2022
03/16/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date02/25/2022
03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - LOCKING/SET SCREWS: SYMPHONY; UNK - RODS : SYMPHONY
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-