• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE X  EP SYSTEM DISPLAY WORKSTATION; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE X  EP SYSTEM DISPLAY WORKSTATION; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number ENSITE-DWS-01
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the atrial fibrillation ablation procedure, a clinically significant delay occurred.Initially the geometry was too large so that during the ablation, without a contact pressure measure, it was not clear when the physician was on the wall of the heart or not with the catheter.The lesion was placed by the system far outside the geometry and this resulted in a clinically significant delay.Despite the issue, the procedure was still able to be completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Additional information: g3, h2, h3, h6.Based off of the information received from the field and review of the ensite x cardiac mapping system verifying that in some cases a model bloat/false space can occur, the reported event was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE X  EP SYSTEM DISPLAY WORKSTATION
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13288632
Report Number2184149-2022-00005
Device Sequence Number14529935
Product Code DQK
UDI-Device Identifier05415067032171
UDI-Public(01)05415067032171(10)7785048
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K202066
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberENSITE-DWS-01
Device Catalogue NumberENSITE-DWS-01
Device Lot Number7785048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-