• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 STRAIGHT SHELL IMPACTOR; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 STRAIGHT SHELL IMPACTOR; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71362299
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a thr surgery, a reflection cup positioner/impactor broke near handle.It is unknown if this happened internally or externally to the patient or if any pieces fell inside of the patient.It is unknown how the procedure was completed or if there was any delay.
 
Additional Manufacturer Narrative
Internal reference number h11.
 
Event or Problem Description
It was reported that, during a thr surgery, an r3 straight shell impactor broke near handle.It is unknown if this happened internally or externally to the patient or if any pieces fell inside of the patient.It is unknown how the procedure was completed or if there was any delay.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be fractured along the welded joint, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are prior actions related to this device and failure mode.The original supplier did not meet the design control specification; therefore, a new supplier is to be selected to minimize the occurrence rate of this failure.However, the likelihood of failure is present as this device is designed to endure multiple surgeries.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 STRAIGHT SHELL IMPACTOR
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13289101
Report Number1020279-2022-00266
Device Sequence Number4547556
Product Code LXH
UDI-Device Identifier03596010443236
UDI-Public03596010443236
Combination Product (Y/N)N
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/18/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71362299
Device Catalogue Number71364450
Device Lot Number13BSA0002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer01/19/2022
02/17/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date01/20/2022
02/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/28/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-