• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. ULTRASITE®; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. ULTRASITE®; SET, ADMINISTRATION, INTRA Back to Search Results
Model Number US1921
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: detailed inquiry description: chemo leak below ultrasite valve on tubing at junction.Leakage noticed after 20 ml of chemo infused.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Two (2) photos were provided for evaluation.A visual evaluation was performed on the photos.The photos showed that the luer was broken off inside of the most distal caresite valve.Although the exact root cause cannot be determined, previous engineering testing cited a probable root cause could be alcohol exposure of certain male adapters, which can lead to difficulty removing or breakage of male luer tapers when connected to female luers.Based on these findings the reported defect was not confirmed to be a manufacturing defect.The end user will be referred to the ifu which states, "swab top lad vigorously for 15 seconds with 70% isopropyl alcohol and allow to air dry prior to use." please note the air drying of the alcohol prior to use.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASITE®
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key13289464
Report Number2523676-2022-00002
Device Sequence Number14555286
Product Code FPA
UDI-Device Identifier04046964187971
UDI-Public(01)04046964187971
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K955585
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUS1921
Device Catalogue NumberUS1921
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer01/05/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date02/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1