Model Number 512538MPA1
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2019
Type of Reportable Event Malfunction
Event or Problem Description
The account alleges that during a balloon-occluded retrograde transvenous obliteration (brto) procedure, for the treatment of gastric varices, the diagnostic catheter tip detached.No medical or surgical intervention was necessary for removal of the detached tip.The tip remained on the guidewire and was removed together from the patient.No additional consequences to report.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect device was returned for evaluation.The complaint was confirmed and the root cause is attributed to clinical use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED.CATHETER
Common Device Name
PERIPHERAL CATHETERS
Manufacturer (Section D)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key13289960
Report Number3010665433-2021-00068
Device Sequence Number19983094
Product Code DQO
UDI-Device Identifier00884450035894
UDI-Public884450035894
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number512538MPA1
Device Lot NumberE1565018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer01/12/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date01/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE.
Patient SexUnknown
-
-