• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® URETHRAL CATHETER ROBINSON MODEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® URETHRAL CATHETER ROBINSON MODEL Back to Search Results
Model Number 056110
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the loose particulates was coming off from the packaging of robinson intermittent catheters.Per additional information received via email on 10jan2022, it was reported that the loose particulates were found by the medline production team.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the loose particulates was coming off from the packaging of robinson intermittent catheters.Per additional information received via email on 10jan2022, it was reported that the loose particulates were found by the medline production team.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.3 out of the 11 samples and 3 out of the 48 samples were confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for patient treatment.The product caused the reported failure.Visual evaluation of the returned sample noted 11 unopened (with original packaging), urethral catheters and 48 additional urethral catheters samples.Visual inspection of the sample noted flaking of the strip outside packaging when touching the packaging, flaking on the gloves and table.Additionally tested 3 out of the 11 samples and 3 out of the 48 samples and flaking immediately occurred when touching packing.Also noted no flaking was observed on the inside of the sealed packaging on the catheter.This is out of specification per inspection procedure, which states "loose foreign matter, engraved and spots >1/32¿ is not allowed".A potential root cause for this failure mode could be ¿inadequate assembly process.Stress on materials during shipment and handling¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review was not performed because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the loose particulates was coming off from the packaging of robinson intermittent catheters.Per additional information received via email on 10jan2022, it was reported that the loose particulates were found by the medline production team.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® URETHRAL CATHETER ROBINSON MODEL
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13290586
Report Number1018233-2022-00103
Device Sequence Number11943951
Product Code KOD
UDI-Device Identifier00801741033506
UDI-Public(01)00801741033506
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/31/2026
Device Model Number056110
Device Catalogue Number056110
Device Lot NumberNGFW2958
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer08/25/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date08/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-