MERIT MEDICAL MANUFACTURING .IMPRESS®.BRAIDE CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 56535RIM
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device will not be returning for evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.
 
Event or Problem Description
The account alleges that during a transcatheter aortic valve replacement (tavr) procedure, the catheter tip detached within the patient.The physician had acquired retrograde femoral artery access and during over-the-wire catheter manipulations within the patients under fluoroscopy, the catheter tip detached.The physician used a vascular snare device to successfully externalized the foreign body liberating the vessel.The patient tolerated the procedure well and was discharged with no additional consequences to report.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
.IMPRESS®.BRAIDE CATHETER
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
8012084662
MDR Report Key13290625
Report Number3010665433-2022-00008
Device Sequence Number4547605
Product Code DQO
UDI-Device Identifier00884450038185
UDI-Public884450038185
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 12/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2024
Device Catalogue Number56535RIM
Device Lot NumberE2089716
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2021
Supplement Date Received by Manufacturer02/23/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INTERVENTIONAL GUIDEWIRE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-