| Catalog Number |
56535RIM |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device will not be returning for evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a transcatheter aortic valve replacement (tavr) procedure, the catheter tip detached within the patient.The physician had acquired retrograde femoral artery access and during over-the-wire catheter manipulations within the patients under fluoroscopy, the catheter tip detached.The physician used a vascular snare device to successfully externalized the foreign body liberating the vessel.The patient tolerated the procedure well and was discharged with no additional consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|