Model Number 350-11-01
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.No information provided in the following section(s).
 
Event or Problem Description
It was reported that during impaction of the left ankle, the surgeon was trying to further seat the tibial just a bit more after already impacting it several times and a small part of the rim of the implant where the poly inserts into broke off.This prevented a trial poly from being placed in the component and ultimately the surgeon took that tibial tray out and put a new one in its place.There was no surgical delay.Patient was last known to be in stable condition following the event.No other patient information or medical history was reported.Devices will be returning.
 
Additional Manufacturer Narrative
Pending evaluation.No information provided in the following section(s).
 
Event or Problem Description
It was reported that during impaction of the left ankle, the surgeon was trying to further seat the tibial just a bit more after already impacting it several times and a small part of the rim of the implant where the poly inserts into broke off.This prevented a trial poly from being placed in the component and ultimately the surgeon took that tibial tray out and put a new one in its place.There was no surgical delay.Patient was last known to be in stable condition following the event.No other patient information or medical history was reported.Devices will be returning.
 
Additional Manufacturer Narrative
Section h10: (h3) the prosthesis fracture reported was likely the result of not using the tibial protector as instructed in the operative technique when impacting the tibial plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANTAGE
Common Device Name
TIBIAL PLATE FB SZ 1 LT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key13290666
Report Number1038671-2022-00030
Device Sequence Number11864385
Product Code HSN
UDI-Device Identifier10885862276131
UDI-Public10885862276131
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K152217
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number350-11-01
Device Catalogue Number350-11-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/30/2021
Supplement Date Received by Manufacturer01/24/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date02/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age65 YR
Patient SexFemale
-
-