| Model Number |
CD3265-40 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
|
| |
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.There were no patient consequences and the patient was in stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Premature depletion was not confirmed by analysis.An abnormal transient battery voltage drop was detected.The voltage drop is consistent with li deposit in the battery.
|
| |
|
Search Alerts/Recalls
|