|
WILLIAM COOK EUROPE NAVALIGN COOK CELECT JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
| Model Number |
G52909 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Structural Problem (2506); Device Tipped Over (2589); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Anxiety (2328); Discomfort (2330); Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It is alleged that the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2013, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown or unavailable.Non-healthcare professional investigation it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged celect is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: investigation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: organ/vena cava perforation, unable to remove, tilt, pain, anxiety, discomfort, limited physical activity.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up.The risks/benefits of filter retrieval should be considered for each patient during follow-up.Once protection from pe is no longer necessary, filter retrieval should be considered.Filter retrieval should be attempted when feasible and clinically indicated.Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e.G., a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications).For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set.It may also be retrieved with the cloversnare® vascular retriever.Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques.The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position.The safety or effectiveness of these alternative retrieval techniques has not been established.Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve.For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Unknown if the reported pain, anxiety, discomfort, limited physical activity is directly related to the filter and unable to identify a corresponding failure mode at this point in time.(b)(4) in lot.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Patient allegedly received an implant on (b)(6) 2013 due to blood clots.Patient is alleging tilt and vena cava/organ perforation, and device is unable to be retrieved.The patient further alleges pain, discomfort, anxiety, and limited physical activity.
|
| |
|
Event or Problem Description
|
|
Patient allegedly received the celect filter via the right internal jugular vein.(b)(6) 2020, per a report from computed tomography; ¿stable appearance of ivc filter with one of the legs either abutting or just protruding into the right wall of the aorta.This finding is unchanged dating back to 2017.¿ (b)(6) 2020, per a report from computed tomography 2; ¿cook celect ivc filter with: 4 mm psoas muscle, 7 mm vertebral, 5 mm aortic, and 6 mm mesenteric perforation.Tilting with the apex against the ivc wall.No fracture, stenosis, or migration.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: b5, b6, annex g, annex b, annex c, annex d, and h6.Investigation: the following allegations have been investigated: aorta perforation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The additional information regarding aorta perforation does not change the previous investigation results for organ/vena cava perforation.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|
|
|