• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH 3202674-005 REDIPAK 1P 10B 50C; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH 3202674-005 REDIPAK 1P 10B 50C; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number PM20022
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the electrode exhibited delamination.Per additional information received, the delamination occurred before use on the patient and there was no impact.
 
Additional Manufacturer Narrative
The device history record (dhr) for lot 005523 was reviewed for both the finished good and the subassemblies manufactured at the plant, and no abnormal process conditions were present during the manufacturing of the products that could have led to the reported condition.The dhr review shows that all acceptance criteria inspections were within acceptable limits during the production process.This includes testing where the liner is peeled from the gel body electrode and inspected for delamination.This product was manufactured between 4-jul-2020 and 10-jul-2020.A sample was received in the form of a photo.The photo shows the open product and the product pouch.Review of the photo shows gel delamination in the upper middle of the pad.The gel delamination does not cross over the carbon mat area where the wires connect to the pad.Gel delamination occurs when the gel-to-release liner bond strength is greater than the cohesive strength of the gel or greater than the gel-to-substrate bond strength.The difference in bond strength can cause the gel to peel from the substrate and/or tear.Delamination is categorized as follows: an aesthetic delamination defect is characterized by a permanent separation of the hydrogel from the substrate, exposing the underlying carbon vinyl with little (i.E., area less than 1/8-inch x 1/8 inch) or no hydrogel remains on the release liner.Also, no silver/silver chloride ink is exposed.Aesthetic delamination defects will not significantly affect electrode function but may displease the clinician.Functional delamination is a permanent separation of the hydrogel from the substrate such that the silver/silver chloride ink remains exposed.Additionally, per our procedure, there shall be no less than 55% hydrogel remaining on the adult electrode when the release liner is removed from the electrode.There shall be no less than 78% hydrogel remaining on the adult electrode, when the area of separation includes the area of underlying conductive mat.With loss of the hydrogel no more than the percentages as defined above the electrodes retain functionally essential performance; however, the reduction in the hydrogel area with the silver/silver chloride ink being exposed may cause an increase in current density across the remaining gel area.This increases the potential for skin irritation and burns.From a root cause analysis perspective, it was not possible to determine a specific root cause.As indicated on the product packaging proper storage and usage of the electrodes is critical to the performance of the gel.The electrodes should be stored in their sealed protective pouch in a cool, dry place and out of direct sunlight.Do not open package until ready for use.Do not bend, fold, or puncture the packaging.Improper storage conditions and/or handling of the product may compromise gel properties prior to the expiration.Due to not being able to verify that the cause of the delamination is manufacturing related there will be no corrective/preventive actions at this time.We will continue to trend this issue for future occurrences as part of the complaint review process.
 
Additional Manufacturer Narrative
Updated section d2 product code from dro to mkj.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3202674-005 REDIPAK 1P 10B 50C
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
CARDINAL HEALTH
2 ludlow parkway
chicopee MA 01022
Manufacturer (Section G)
CARDINAL HEALTH
2 ludlow parkway
chicopee MA 01022
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183840
MDR Report Key13294243
Report Number1219103-2022-00345
Device Sequence Number17280034
Product Code MKJ
UDI-Device Identifier00681490580656
UDI-Public00681490580656
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/09/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberPM20022
Device Catalogue NumberPM20022
Device Lot Number005523
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer01/04/2022
01/04/2022
01/04/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date02/16/2022
05/17/2022
08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-