U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1040-080
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a mildly calcified 70% stenosed lesion in the atrioventricular fistula.A 4x80mm armada 35 balloon dilatation catheter (bdc) was advanced into the fistula and inflated one time to 16 atmospheres (atm) when the balloon ruptured.The bdc was removed, and another armada 35 was used to successfully complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13297865
Report Number2024168-2022-00673
Device Sequence Number12006448
Product Code LIT
UDI-Device Identifier08717648154539
UDI-Public08717648154539
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberB1040-080
Device Catalogue NumberB1040-080
Device Lot Number91217G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/28/2021
Supplement Date Received by Manufacturer02/16/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date03/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-