• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW ADULT NASAL CANNULA; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW ADULT NASAL CANNULA; CAT Back to Search Results
Model Number OPT844
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2021
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare facility in (b)(6) reported that the tubing of an opt844 optiflow adult nasal cannula was found damaged before patient use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of investigation.
 
Event or Problem Description
A healthcare facility in china reported that the tubing of an opt844 optiflow adult nasal cannula was found damaged before patient use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The opt844 interface is used to deliver humidified oxygen to patients.The opt844 consists of a lightweight delivery tube which is connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's nares.Method: the complaint (b)(4) nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information provided by the customer, previous investigations of similar complaints, and our knowledge of the product results: the customer has stated that the tubing of an opt844 optiflow adult nasal cannula was found damaged.The customer further confirmed that the interface was found damaged.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.However, the cause of the reported event is most likely due to overtightening of the head strap.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is disposed of.The subject cannula would have met the specification at the time of production.The user instructions which accompany theopt844 optiflow adult nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "do not crush or stretch tube.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW ADULT NASAL CANNULA
Common Device Name
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534002
MDR Report Key13298826
Report Number9611451-2022-00073
Device Sequence Number20115035
Product Code CAT
UDI-Device Identifier09420012432124
UDI-Public(01)09420012432124(10)2100206868(11)170501
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT844
Device Catalogue NumberOPT844
Device Lot Number2100206868
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer01/24/2022
Initial Report FDA Received Date01/19/2022
Supplement Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-