STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1896-5045S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369); Implant Pain (4561)
Date of Event 09/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition is unknown.
 
Event or Problem Description
Customer reported a broken gamma nail 3 titanium.On the (b)(6) 2021 the broken gamma nail was removed and a pseudarthrosis treatment was performed (tissue sampling, cleaning of fracture site, re-osteosynthesis with a proximal femoral plate on the lateral side and an ao small fragment plate on the ventral side with spongiosa/filling with bone- chips).A broken screw was also reported.
 
Additional Manufacturer Narrative
Please note the device was returned, and h6 (method, results and conclusion codes).The reported event was be confirmed.Review of dhr, complaint history, capa databases and risk analysis did not identify any discrepancies.No nc/capa had been filed for the item in question.Potential nail breakage had been considered in the risk management file.Potential reasons for nail breakage are listed in the labelling.The screw was completely broken in the approx.In the middle of the shaft.The screw had broken in fatigue manner while positioned in dynamic mode in the oblong drill hole at distal.The cause of the event was determined being exceeding axial load application on the whole construct of implants which was caused by patient and suboptimal fragment reduction by the user.In case further substantive information becomes available we reserve the right to reopen the investigation with root cause assessment.
 
Event or Problem Description
Customer reported a broken gamma nail 3 titanium.On the (b)(6) 2021 the broken gamma nail was removed and a pseudarthrosis treatment was performed (tissue sampling, cleaning of fracture site, re-osteosynthesis with a proximal femoral plate on the lateral side and an ao small fragment plate on the ventral side with spongiosa/filling with bone- chips) a broken screw was also reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13303789
Report Number0009610622-2022-00029
Device Sequence Number14686935
Product Code HSB
UDI-Device Identifier04546540202505
UDI-Public04546540202505
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2026
Device Model Number1896-5045S
Device Catalogue Number18965045S
Device Lot NumberK07B8DE
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/24/2021
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date01/20/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight95 KG
-
-