Model Number 9733858 |
Device Problems
Mechanical Problem (1384); Device Sensing Problem (2917); Human-Device Interface Problem (2949)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess).It was reported that the non-invasive patient tracker (nipt) was missing.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no reported impact on patient outcome.Additional information was received stating that after some changes done by the site, the patient tracker wasn't recognized anymore by the system in the ent profile.There wasn't any physical unit missing.Hence, without the patient tracker it wasn't possible to use the system in the ent profile.There was no reported impact to the patient.
|
|
Manufacturer Narrative
|
H2) additional information was added to the event description.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received stating that the procedure was completed under the neuro profile and not the ent (due to the missing tracker in the ent profile).Navigation was not aborted.Troubleshooting was not performed at the time of the event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received.The system does not work anymore.The screen displayed no signal.On site, there was an old computer so they made a switch and they concluded that it was the computer.
|
|