MEDOS INTERNATIONAL SÃ RL CH CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES; NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE
|
Back to Search Results |
|
| Model Number |
283903611 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021 during a lumbar arthrodesis fixation surgery, the confidence spinal cement needle broke.Part of the needle remained inside the bone (vertebra) without causing damage or injury.There is no further information available.This report is for one (1) confidence spinal cement system introducer needle, diamond tip 11g x 6 inches.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A manufacturing record evaluation was performed for the finished device.Product code: 283903611.Lot number: 324580.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: september 30, 2021.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the photos.Visual analysis of the provided photo found the conf intro ndle, dia, 11g 6" appears to be bent as well as the tip has broken off.No evidence of the broken fragment being left inside the patient was provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for conf intro ndle, dia, 11g 6".There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H6: medical device problem code.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|