Model Number 338.26
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on an unknown date in 2021, the tip for dhs impactor was broken.A fragment was generated and was easily removed without additional intervention.There was a surgical delay of two (2) minutes.The procedure was successfully completed.There was no patient consequence.This report is for one (1) tip for dhs®/dcs® impactor (338.28).This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Date of event: event occurred on an unknown date in 2021.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Service and repair evaluation: the customer reported the 338.26, tip for dhs/dcs impactor (338.28), lot#: h786340, dhs impactor cap is broken.The repair technician reported, the handle of impactor is damaged broken and pieces are missing.Handle cracked/broken is the reason for repair.The item is not repairable, per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.H4 device history: part #: 338.26, synthes lot #: h786340, supplier lot #: h786340, release to warehouse date: 10/15/2019, supplier: (b)(4).No ncrs were generated, during production.Review of the device history record(s) showed, that there were no issues, during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained, that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIP FOR DHS®/DCS® IMPACTOR (338.28)
Common Device Name
IMPACTOR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13308337
Report Number2939274-2022-00228
Device Sequence Number14308842
Product Code HWA
UDI-Device Identifier10886982193292
UDI-Public(01)10886982193292
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number338.26
Device Catalogue Number338.26
Device Lot NumberH786340
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/26/2021
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date01/20/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-