• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 4.5MM/6.5MM STEPPED DRILL BIT LARGE QC/485MM; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 4.5MM/6.5MM STEPPED DRILL BIT LARGE QC/485MM; BIT, DRILL Back to Search Results
Catalog Number 03.010.078
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the drill bits were damaged during the procedure.Two of the broken drill bits fragments remain in the patient¿s bone.Two of the hip screws were implanted outside of the nail lfn.The screws were extracted and there was a forty (40) minutes surgical delay.The surgery was completed successfully.This report involves one (1) 4.5mm/6.5mm stepped drill bit large qc/485mm.This is report 4 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9, h3, h6: the device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: product code #: 03.010.078, synthes lot #: f-18810, manufacturing site: werk villmergen logistik, release to warehouse date: march 22, 2016, supplier: (b)(4).A manufacturing record evaluation was performed for the non-sterile article lot and no non-conformances were identified.Investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the 4.5/6.5 stepped drill bit large qc/485 was found broken from the tip of the drill; the broken part was not returned for investigation.A dimensional inspection was unable to be performed due to post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 4.5/6.5 stepped drill bit large qc/485 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This is report 5 of 5 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a photo investigation was completed: the device was not returned for visual inspection.The images provided have been reviewed, complaint is not confirmed.After reviewing the given information, is not possible to determine any device related problem.A definite assignable root cause could not be determined based on the provided information.As the device was not returned, and as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM/6.5MM STEPPED DRILL BIT LARGE QC/485MM
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13308477
Report Number8030965-2022-00446
Device Sequence Number4978450
Product Code HTW
UDI-Device Identifier07611819765702
UDI-Public(01)07611819765702
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.078
Device Lot NumberF-18810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/22/2021
Supplement Date Received by Manufacturer01/14/2022
02/11/2022
03/15/2022
Initial Report FDA Received Date01/20/2022
Supplement Report FDA Received Date02/11/2022
02/24/2022
03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRILL BIT Ø4.2 CALIBR L145 3FLUTE; DRILL BIT Ø4.2 CALIBR L145 3FLUTE; DRILL BIT Ø4.2 CALIBR L145 3FLUTE; DRILL BIT Ø4.2 CALIBR L145 3FLUTE; HIP SCR Ø6.5 SELF-TAP L90 F/EXPERT LFN T; HIP SCR Ø6.5 SELF-TAP L90 F/EXPERT LFN T; HIP SCR Ø6.5 SELF-TAP L95 F/EXPERT LFN T; HIP SCR Ø6.5 SELF-TAP L95 F/EXPERT LFN T
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-