MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRUETOME JAG 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRUETOME JAG 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00583140
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a truetome jag 44 was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, after the physician inserted the jagwire guidewire into the cbd, the truetome was connected to the generator for sphincterotomy.However, during sphincterotomy, the cutting wire was broken.It was reported that the wire anchor had dislodged from the catheter.Additionally, no part of the cutting wire detached and fell into the patient.The physician removed the device and the procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Event or Problem Description
It was reported to boston scientific corporation that a truetome jag 44 was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, after the physician inserted the jagwire guidewire into the cbd, the truetome was connected to the generator for sphincterotomy.However, during sphincterotomy, the cutting wire was broken.It was reported that the wire anchor had dislodged from the catheter.Additionally, no part of the cutting wire detached and fell into the patient.The physician removed the device and the procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6 (device codes): medical device problem code a051201 captures the reportable event of wire anchor dislodged.Block h10: the returned truetome jag 44 was analyzed, and a visual evaluation noted that the cutting wire anchor was dislodged/dislocated from the distal pierced hole, consistent with the findings when the device was observed under magnification.Additionally, the cutting wire was blackened, and the working length was also torn from the distal pierced hole.No other problems with the device were noted.The reported event of cutting wire anchor dislodged was confirmed.Upon analysis, it was found that the cutting wire anchor was dislodged/dislocated.The cutting wire was blackened with no evidence of a broken or missing wire.The cutting wire being blackened indicates that the device was energized.The working length was also torn at the distal pierced hole and this could have been caused by submitting the cutting wire to tension during the handle actuation.In addition, with the possibility of the device being energized during the handle actuation, the analyzed problem could happen.Also, bowing the device without being completely out of the scope can lead to a tear in the working length, displacing the cutting wire anchor from its position.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETOME JAG 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13308854
Report Number3005099803-2022-00141
Device Sequence Number3424171
Product Code KNS
UDI-Device Identifier08714729858683
UDI-Public08714729858683
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K122203
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/27/2023
Device Model NumberM00583140
Device Catalogue Number8314
Device Lot Number0027396441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/27/2021
Supplement Date Received by Manufacturer02/11/2022
Initial Report FDA Received Date01/20/2022
Supplement Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
Patient Weight77 KG
-
-