| Model Number |
FLX |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The service personnel inspected the modules and checked the alignment of the no stop divert devices at the entry of the internal u-turn modules.The involved modules are kept monitored for further investigation.
|
| |
|
Event or Problem Description
|
|
Cases of sample spillage have been reported from the field at the entry of some internal u-turn modules.This module provides a short way for carriers to reach their destination on the automation system by diverting them to a u-shaped lane connecting a portion of track with opposite flow direction.The sample spillage may cause cross contamination if the sample drops fall into another uncapped sample tube present in the queue or in the main lane.Sample residues have been found on the track profile.The investigation performed up to now did not provide any evidence of cross contamination.No discrepant test results potentially caused by sample cross contamination have been identified.
|
| |
|
Search Alerts/Recalls
|