DEPUY ORTHOPAEDICS INC US HP MBT DRILL TOWER; DRILLS (BITS, SLEEVES, GUIDES) : DRILL GUIDES Back to Search Results
Model Number 9505-02-027
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon performed a total knee arthroplasty using the lcs total knee system.During post operative radiography a small piece of metal was detected.Surgeon decided to perform an exploration of the knee and the item was removed.It was later determined to be a prong from the mbt drill tower.All pieces were accounted for.Surgical delay of 25 min occurred.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon performed a total knee arthroplasty using the lcs total knee system.During post operative radiography a small piece of metal was detected.Surgeon decided to perform an exploration of the knee and the item was removed.It was later determined to be a prong from the mbt drill tower.All pieces were accounted for.Surgical delay of 25 min occurred.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary = the device associated with this report was returned for analysis.Visual evaluation of the product confirmed the complaint.One of the prongs broke off from the hp mbt drill tower.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot = the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HP MBT DRILL TOWER
Common Device Name
DRILLS (BITS, SLEEVES, GUIDES) : DRILL GUIDES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13309520
Report Number1818910-2022-01299
Device Sequence Number14255062
Product Code HWT
UDI-Device Identifier10603295227472
UDI-Public10603295227472
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9505-02-027
Device Catalogue Number950502027
Device Lot NumberB0208
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer02/25/2022
03/14/2022
05/02/2022
Initial Report FDA Received Date01/20/2022
Supplement Report FDA Received Date03/01/2022
03/17/2022
05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-