Model Number 6R45B
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Component code: (b)(4).Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the 6r45b reload was received with the reload lock out damaged and partially fired 1/4.No functional test could be performed due to the condition of the reload.The event reported was confirmed and it is related to improper use of the device.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If the re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during an unknown procedure, the device could not be fired at the 1st firing.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELOADS FOR LINEAR CUTTERS
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
*  
6107428552
MDR Report Key13309606
Report Number3005075853-2022-00405
Device Sequence Number3424714
Product Code GDW
UDI-Device Identifier10705036000846
UDI-Public10705036000846
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/30/2025
Device Model Number6R45B
Device Catalogue Number6R45B
Device Lot NumberU5F702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/22/2021
Initial Report FDA Received Date01/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-