| Model Number |
L151-596 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the tip of the prolift inserter fractured while disengaging from a cage intra-operatively.
|
| |
|
Additional Manufacturer Narrative
|
|
Device not returned it is hypothesized that during use an excessive force was applied to the installer at the angle which resulted in a torsional or off-ais asymmetric (rocking or toggle) load being applied to the distal tip of the installer.An expandable spacer installers are not designed to withstand an excessive torsional or off-axis load, an insertion forces applied at an angle may have resulted in the failure at the distal tip of the installer.
|
| |
|