ALLERGAN (COSTA RICA) STYLE 110 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 110-150
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Seroma (2069)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.The events of "seroma and capsular contracture" are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: seroma and capsular contracture.
 
Event or Problem Description
Patient reported right side rupture.Healthcare professional reported ¿breast cancer¿, ¿capsular contracture of breast implant¿, ¿slightly increased, small amount of fluid collection around the implant bag¿, ¿prominent internal folding¿.The event of ¿malignant neoplasm¿ is not device related.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: opening in the shell, underweight, crease flat and wear abrasion.A microscopic analysis was performed which identify a sharp edge opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is an opening sharp in the posterior side assessed as unidentified (tear) broken.
 
Event or Problem Description
Patient reported right side rupture.Healthcare professional reported ¿breast cancer¿, ¿capsular contracture of breast implant¿, ¿slightly increased, small amount of fluid collection around the implant bag¿, ¿prominent internal folding¿.The event of ¿malignant neoplasm¿ is not device related.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 110 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key13309817
Report Number9617229-2021-60241
Device Sequence Number11599770
Product Code FTR
UDI-Device Identifier10888628002173
UDI-Public10888628002173
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/15/2023
Device Model Number110-150
Device Catalogue Number110-150
Device Lot Number3167370
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer01/27/2022
Initial Report FDA Received Date01/20/2022
Supplement Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-