Based on the investigation analysis, the event that was reported on (b)(6) 2022 at 10:40 am pst (pacific standard time, user's local time) was on the day after insertion and the system was in initialization phase.At 1:41 am est (eastern standard time), the system displayed a sensor reading of 45 mg/dl.The reported fingerstick measurement of 91 mg/dl was not entered in the system as a calibration entry.Per the reported event, there was temporary mismatch between the sensor readings and fingerstick measurements.The hypo alert was asserted before the reported event.Further analysis of the sensor diagnostic data showed the sensor was settling after insertion and this caused some transient noise in the optical channel measurements.Once the sensor settled, the system displayed better agreement between the sensor readings and the fingerstick measurements.User didn't mention if he was symptomatic or if he solved the event by himself, but didn't require medical treatment or hospitalization.The system is displaying better agreement between the sensor readings and the fingerstick measurements.No further resolution was necessary for this complaint.H6.Investigation findings updated to 213.H6.Investigation conclusions updated to 67.
|