SYNTHES GMBH AIR OSCILLATOR; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
511.610 |
| Medical Device Problem Codes |
Unintended System Motion (1430); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported from russia that during service and evaluation, it was determined that the air oscillator device had unintended activation/motion, and corrosion/rusting/pitting.It was further determined that the device failed pretest for check for unintended motion.It was noted in the service order that the equipment did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.The air oscillator device was evaluated and the reported condition that the equipment was not working was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had unintended motion.The assignable root cause of this condition was determined to be traced to the user, which is user error.Udi ¿ (b)(4).
|
| |
|