| Model Number |
BSM-6501A |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Technical support (ts) suggested for the customer to check some things on the wlan box attached to the bsm.Ts suggested for him to change out the wlan box with a known working one.The customer will try these suggestion and reach out again if needed.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the bedside monitor (bsm) is having intermittent communication loss (comm loss).There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: the customer reported that the bedside monitor (bsm) was in intermittent comm loss at the central nurse's station (cns).Technical support (ts) suggested the customer change out the wlan box with a known working one.The customer was to try this and reach out again if needed.No patient harm was reported.Investigation summary: the customer checked to verify that the physical connection between the wlan pack and the bsm 6000 was seated properly.On 2/23/2022, the customer stated that they resolved the issue by resetting the wlan pack.The wireless lan station is an optional module that is connected to the bedside device as.Since the cause of the reported event was isolated to the setting of the wlan pack, service history for this facility was reviewed for issues related to qi-420p.Additionally, the service history was reviewed for the three-month period between 11/2021 to 01/2022 for related incidents.This was an isolated incident.There has been no recurrence of communication loss related to wlan.As a result, the probable cause is attributable to configuration and/or network related.There is no evidence of a qi-420p, bsm 6000, or cns malfunction.Comm loss/signal loss is an error message notifying the user of loss of network communication between the monitoring cns or rns and that of the monitored devices.The message appears on an individual "tile" on the cns, corresponding to the communication loss of the monitored device.Other related events that may cause the issue are accidentally turning off the device, or accidentally unplugging the network cable of the monitored device, or user error.Additional device information: d10 concomitant medical device: the following devices were used in conjunction with the cns: bsm: model #: ni.Serial #: ni.Wlan device: model #: ni.Serial #: ni.
|
| |
|
Event or Problem Description
|
|
The customer reported that the bedside monitor (bsm) was in intermittent comm loss at the central nurse's station (cns).There was no patient injury reported.
|
| |
|