| Model Number |
G47449 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer ref# (b)(4).Pma/510(k): p180001.Investigation is still in progress.
|
| |
|
Event or Problem Description
|
|
Description of event: according to the initial reporter: a patient of undisclosed gender and age underwent a tvar procedure in which the zenith tx2 dissection endovascular graft straight component, g47449, was used.The trigger wire knob was unable to be released.The physician had to use excessive force and pulled 4-5 times with such force with small breaks in between from pulling so hard.The physician stated that in using only these products in fifteen years this is the hardest he has ever had to pull of any device.It was noted there was hardly any tortuous anatomy, no calcification, and access was good.The physician was eventually able to release the graft component.Patient outcome: according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.The patient did not require any additional procedures due to this occurrence.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer ref# (b)(4).Summary of investigational findings: it was reported by phone that a patient of undisclosed gender and age underwent a tevar procedure in which the zdeg-p-32-202-pf-us (complaint device) was used.It was noted there was hardly any tortuous anatomy, no calcification, and access was good.It was reported that the trigger wire knob was unable to be released.The physician had to use excessive force and pulled 4-5 times with such force with small breaks in between from pulling so hard.The physician was eventually able to release the graft component.No harm to patient was reported.Review of the device history record gave no indication of the device being produced out of specification.Per the instructions for use sent with the device, the green trigger wire knob should be withdrawn in a continuous movement until the proximal end of the graft opens.The green trigger wire knob should not be rotated.This complaint is based on the provided information from customer.No imaging was available, or product was returned.Based on the provided information, it was not be possible to determine the cause of the reported failure.However, unintended user error could possibly contribute to the reported event.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|