STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-2530
Medical Device Problem Codes Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Discomfort (2330); Metal Related Pathology (4530)
Date of Event 04/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned to the manufacturer.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on her left hip on or about (b)(6) 2012 and was revised on (b)(6) 2021.It is further alleged that she suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in her blood.
 
Additional Manufacturer Narrative
Reported event: an event regarding fretting and altr/abnormal ion level involving an accolade stem that was mated with an lfit v40 cocr head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Clinician review: a review of the provided medical information by a clinical consultant indicated: "revision of a 40 mm lfit v40 femoral head to a ceramic 36mm with matching liner is confirmed.Findings in the operative report for this surgery describe metallosis and tissue changes consistent with altr.In the indications section of the operative report the surgeon states there were "clinical and serologic" evidence of altr.However, there is no laboratory reports confirming elevated serum metal ions.The root cause of the trunnionosis and altr can not be established with the documentation provided." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on her left hip on or about (b)(6) 2012 and was revised on (b)(6) 2021.It is further alleged that she suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in her blood.Update 04-april-2022: medical review of provided medical records indicated the following: "the revision operative report described a metallosis throughout the hip as well tissue changes consistent with adverse local tissue reaction (altr) upon the removal of the 40 mm lfit v40 femoral head a large amount of black tissue was seen on the trunnion.The black tissue was removed with a scalpel and a sponge.Once cleansed the trunnion was inspected and found to be undamaged.Both the femoral and acetabular components were seen to be well fixed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 2.5 ACCOLADE (127 DEG) SIZE 2.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13313449
Report Number0002249697-2022-00117
Device Sequence Number19212027
Product Code LPH
UDI-Device Identifier04546540510747
UDI-Public04546540510747
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Model Number6021-2530
Device Catalogue Number6021-2530
Device Lot Number38569103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date01/21/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/06/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexFemale
-
-