• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION Back to Search Results
Model Number 8015
Device Problems Device Alarm System (1012); Break (1069); Use of Device Problem (1670); Battery Problem (2885); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Asystole (4442); Insufficient Information (4580)
Event Date 12/26/2021
Event Type  Death  
Manufacturer Narrative
A follow up report will be submitted with investigation results should the device be repaired or the device/logs be received for evaluation.Per 803.52(f)(11)(iii) the information provided represents all of the known information at this time.The complainant or reporter was unable or unwilling to provide any further patient, product, or procedural details to the manufacturer.Device was not returned to manufacturing facility.
 
Event Description
It was reported that the pump module had a channel disconnect alarm then battery failure.It was noted that the pump module had stopped.There was patient involvement but impact was unknown.
 
Manufacturer Narrative
It was determined through investigation of the returned devices that the initially reported suspect device with serial number# (b)(4)is a concomitant and this file has captured the correct suspect device with serial number# (b)(4) as per investigation report.
 
Event Description
It was reported that the pump module had a channel disconnect alarm then battery failure.It was noted that the pump module had stopped.Received a copy of the customer's medwatch report from fda which states "in afternoon, nursing noted a pump malfunction (channel disconnect error) ¿ propofol, levophed, and vasopressin were not running.Epinephrine was running in a separate functioning channel.Patient's mean arterial pressure (map's) dropped.It took approximately 30 seconds for nursing to get in the room and turn the channel pumps to turn back on, which the channels then said "battery error" and the channels shut off.With re-initiation of pump, patient's maps did not improve.The patient continued to desaturate and became bradycardic which then progressed to asystole.Resuscitation was not attempted as the patient was made dnr (do not resuscitate) earlier in the day.Given the patient's clinical condition, it cannot be determined to what effect this hastened the patient's death." it was reported that the primary and secondary causes of death were hypoxic respiratory failure secondary to covid pneumonia.
 
Event Description
It was reported that the pump module had a channel disconnect alarm then battery failure.It was noted that the pump module had stopped.Received a copy of the customer's medwatch report from fda which states "in afternoon, nursing noted a pump malfunction (channel disconnect error) ¿ propofol, levophed, and vasopressin were not running.Epinephrine was running in a separate functioning channel.Patient's mean arterial pressure (map's) dropped.It took approximately 30 seconds for nursing to get in the room and turn the channel pumps to turn back on, which the channels then said "battery error" and the channels shut off.With re-initiation of pump, patient's maps did not improve.The patient continued to desaturate and became bradycardic which then progressed to asystole.Resuscitation was not attempted as the patient was made dnr (do not resuscitate) earlier in the day.Given the patient's clinical condition, it cannot be determined to what effect this hastened the patient's death." it was reported that the primary and secondary causes of death were hypoxic respiratory failure secondary to covid pneumonia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYSTEM
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
brett wilko
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key13313736
MDR Text Key284173978
Report Number2016493-2022-100366
Device Sequence Number1
Product Code FRN
UDI-Device Identifier10885403801518
UDI-Public(01)10885403801518
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133532
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8015
Device Catalogue Number8015 ALARIS PCU 1.5 MODULE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-1520-2017
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age52 YR
Patient SexFemale
-
-