• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FEM RT SZ 10 CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FEM RT SZ 10 CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Model Number 1504-10-210
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the device associated with this report was received for examination.Review of the returned product found foam packaging material adhered/melted to the surface of the implant.Root cause and corrective action are documented in the capa system.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event or Problem Description
It was reported that last (b)(6) 2021 a size 10 right attune ps cemented femur was defective on the m/l position of the impact.There had a appeared to be some flaky material coated along the ridge of the implant.The defective implant was saved and packaged properly in a hazardous waste bag.Opened a new femur of the same size and kept the defective on off to the side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM RT SZ 10 CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND - 9616671
loughbeg, ringaskiddy co
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13313836
Report Number1818910-2022-01330
Device Sequence Number1561869
Product Code NJL
UDI-Device Identifier10603295041818
UDI-Public10603295041818
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-10-210
Device Catalogue Number150410210
Device Lot Number505129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2022
Initial Report FDA Received Date01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-