Model Number SSL-320
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 12/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade iv.Device has been explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified, weight to the spec and deformation.After autoclave cycle were observed, voids and cloudy color in the gel.Based on the device analysis, the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Healthcare professional reported, left side capsular contracture, baker grade iv.Device has been explanted.
 
Event or Problem Description
Healthcare professional reported, left side capsular contracture, baker grade iv.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key13314005
Report Number9617229-2022-00123
Device Sequence Number13848027
Product Code FTR
UDI-Device Identifier10888628032552
UDI-Public10888628032552
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/21/2024
Device Model NumberSSL-320
Device Catalogue NumberSSL-320
Device Lot Number3391338
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/30/2021
Supplement Date Received by Manufacturer01/26/2022
05/02/2022
Initial Report FDA Received Date01/21/2022
Supplement Report FDA Received Date02/22/2022
05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale