C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH NICORE¿ GUIDEWIRE
|
Back to Search Results |
|
| Model Number |
776424 |
| Medical Device Problem Code |
Calcified (1077)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the user experienced stone formation and stent encrustation, ureteral reflux, infection, and urinary symptoms, occlusion while using the ureteral stent.
|
| |
|
Event or Problem Description
|
|
It was reported that the user experienced stone formation and stent encrustation, ureteral reflux, infection, and urinary symptoms, occlusion while using the ureteral stent.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is inconclusive because no sample was returned.A potential root cause for this event could be, "material selection".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "potential complications associated with retrograde/antegrade positioning of indwelling ureteral stents include the following: edema; stone formation; peritonitis; extravasation; ureteral reflux; stent dislogdgement, fragmentation, migration, occlusion; fistula formation; loss of renal function; hemorrhage; pain/discomfort; stent encrustation; hydronephrosis; perforation of kidney, renal pelvis, ureter and/or bladder; ureteral erosion; infection; urinary symptoms"."ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient¿s condition and other patient specific factors.When long-term use is indicated, it is recommended that indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.".H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|